Journal of Biomedical Engineering ; (6): 329-338, 2022.
Article in Chinese | WPRIM | ID: wpr-928229

ABSTRACT

The implantation of biventricular assist device (BiVAD) is more challenging than that of left ventricular assist device for the interaction in the process of multiple input and output. Besides, ventricular assist device (VAD) often runs in constant speed (CS) mode in clinical use and thus BiVAD also faces the problems of low pulsation and imbalance of blood volume between systemic circulation and pulmonary circulation. In this paper, a delay assist mode for a VAD by shortening the support time of VAD was put forward. Then, the effect of the delay mode on cardiac output, pulsation and the function of the aortic valve was observed by numerical method and the rules of hemodynamics were revealed. The research showed that compared with VAD supported in CS mode, the VAD using delay mode in systolic and diastolic period proposed in this paper could meet the demand of cardiac output perfusion and restore the function of the arterial valves. The open ratio of aortic valve (AV) and pulmonary valve (PV) increased with the time set in delay mode, and the blood through the AV/PV helped to balance the left and the right cardiac volume. Besides, delay mode also improved the pulsation index of arterial blood flow, which is conducive to the recovery of the ventricular pulse function of patients.


Subject(s)
Humans , Cardiovascular System , Diastole , Heart Failure , Heart Rate , Heart-Assist Devices , Hemodynamics , Models, Cardiovascular
SELECTION OF CITATIONS